Skip to main content
SLU publication database (SLUpub)

Abstract

A strategy for optimizing LC-MS metabolomics data processing is proposed. We applied this strategy on the XCMS open source package written in R on both human and plant biology data. The strategy is a sequential design of experiments (DoE) based on a dilution series from a pooled sample and a measure of correlation between diluted concentrations and integrated peak areas. The reliability index metric, used to define peak quality, simultaneously favors reliable peaks and disfavors unreliable peaks using a weighted ratio between peaks with high and low response linearity. DoE optimization resulted in the case studies in more than 57% improvement in the reliability index compared to the use of the default settings. The proposed strategy can be applied to any other data processing software involving parameters to be tuned, e.g., MZmine 2. It can also be fully automated and used as a module in a complete metabolomics data processing pipeline.

Published in

Analytical Chemistry
2012, volume: 84, number: 15, pages: 6869-6876
Publisher: AMER CHEMICAL SOC

SLU Authors

UKÄ Subject classification

Analytical Chemistry

Publication identifier

  • DOI: https://doi.org/10.1021/ac301482k

Permanent link to this page (URI)

https://res.slu.se/id/publ/42596