Abstract

The Sysmex XT-2000iV automated differential leukocyte count performed well with most samples from diseased dogs, cats, and horses. Basophils were not detected. Immature neutrophils or prominent toxic changes often induced errors in samples from cats and dogs.

Keywords

Blood cell; hematology analyzer; laboratory; instrumentation; validation

Published in

Veterinary Clinical Pathology
2009, volume: 38, number: 2, pages: 175-182
Publisher: AMER SOC VETERINARY CLINICAL PATHOLOGY

SLU Authors

UKÄ Subject classification

Veterinary Science
Animal and Dairy Science

Publication identifier

  • DOI: https://doi.org/10.1111/j.1939-165X.2009.00126.x