Skip to main content
SLU publication database (SLUpub)

Abstract

Real-time fluorescence monitoring has been mostly performed in marine systems, with little progress being made in the application of fluorescence excitationeemission matrix (EEM) spectroscopy, especially for freshwater monitoring. This paper presents a two weeks experiment where real-time fluorescence EEM data have been obtained for Bourn Brook, Birmingham, UK, using an in-situ fibre-optic probe. Fluorescence EEMs were measured every 3 min for two weeks, with control ‘grab’ samples every hour analyzed for fluorescence EEMs as well as pH, conductivity and dissolved organic carbon. Comparison of realtime and control samples showed an excellent agreement, with no evidence of fibre-optic probe fouling. EEMs of different character were identified using self-organizing maps, which demonstrated seven clusters of fluorescence EEMs which related to the intensity of fluorescence and relative intensities of peak T1 and T2 vs. peak C and peak A fluorescence. Fluorescence intensity of peaks A and C were observed to increase with rainfall, and a diesel pollution event was detected through an increase in T2 fluorescence.

Keywords

Fluorescence spectroscopy; Water quality monitoring; Dissolved organic matter

Published in

Water Research
2010, volume: 44, number: 18, pages: 5356-5366
Publisher: Elsevier

SLU Authors

UKÄ Subject classification

Environmental Sciences

Publication identifier

  • DOI: https://doi.org/10.1016/j.watres.2010.06.036

Permanent link to this page (URI)

https://res.slu.se/id/publ/78265