Skip to main content
SLU publication database (SLUpub)

Abstract

Objective To compare the maximum pressure sustained by airways sealed with a resorbable ligation device (LigaTie) and with thoracoabdominal (TA) staples during thoracoscopic-assisted hilar lung lobectomies. Study design In situ cadaveric study. Animals Ten feline cadavers. Methods Twenty lung lobectomies were performed with either a LigaTie or a TA stapler (n = 10 per group) by using a thoracoscopic-assisted technique. Procedures were randomly performed on the cranial lung lobes only. The distance between the ligation and the hilus was measured on each specimen. Airway pressures were tested to a maximum pressure of 40 cmH(2)O. Results All LigaTie sites reached a maximum airway pressure without leaking and were easily placed at the hilus. Four of 10 TA stapled sites leaked at a mean pressure of 22 cmH(2)O (P = .010). All leaks occurred when the TA staple line was >5 mm from the hilus; no leaks occurred when the TA staple line was

Published in

Veterinary Surgery
2019, volume: 48, number: 4, pages: 563-569

SLU Authors

UKÄ Subject classification

Surgery

Publication identifier

  • DOI: https://doi.org/10.1111/vsu.13109

Permanent link to this page (URI)

https://res.slu.se/id/publ/97164