Skip to main content
SLU publication database (SLUpub)

Research article2013Peer reviewedOpen access

Focal knee resurfacing and effects of surgical precision on opposing cartilage. A pilot study on 12 sheep

Martinez-Carranza, Nicolas; Berg, Hans E.; Hultenby, Kjell; Nurmi-Sandh, Hilkka; Ryd, Leif; Lagerstedt, Anne-Sofie

Abstract

Conclusions: Our results showed significant surgical imprecision and protruding implants imposed severe cartilage damage. We therefore suggest implants should be placed recessed (approx. 0.5 mm) below the surrounding cartilage in this animal model. These results encourage further studies of metallic implants yet the utmost precision regarding position is required. (C) 2013 Osteoarthritis Research Society International. Published by Elsevier Ltd. All rights reserved.

Keywords

Focal cartilage lesions; Focal knee resurfacing; Metal implants; Animal study

Published in

Osteoarthritis and Cartilage
2013, Volume: 21, number: 5, pages: 739-745
Publisher: ELSEVIER SCI LTD