Lilliehöök, Inger
- Department of Clinical Sciences, Swedish University of Agricultural Sciences
Research article2009Peer reviewed
Lilliehöök, Inger; Tvedten, Harold
The Sysmex XT-2000iV automated differential leukocyte count performed well with most samples from diseased dogs, cats, and horses. Basophils were not detected. Immature neutrophils or prominent toxic changes often induced errors in samples from cats and dogs.
Blood cell; hematology analyzer; laboratory; instrumentation; validation
Veterinary Clinical Pathology
2009, Volume: 38, number: 2, pages: 175-182 Publisher: AMER SOC VETERINARY CLINICAL PATHOLOGY
Veterinary Science
Animal and Dairy Science
DOI: https://doi.org/10.1111/j.1939-165X.2009.00126.x
https://res.slu.se/id/publ/49838