Comparison of UPLC and HPLC for Analysis of Dietary Folates

Jastrebova Jelena, Strandler Hanna Sara, Patring Johan, Wiklund Tony

Abstract

Ultra performance liquid chromatography (UPLC) using small sub-2 mu m particles and high performance liquid chromatography (HPLC) were compared for separation and determination of the most common dietary folates; 5-methyltetrahydrofolate, tetrahydrofolate, 5-formyltetrahydrofolate, 10-formylfolic acid and folic acid. Two UPLC columns-Acquity BEH C(18) and Acquity HSS T3, and two HPLC columns with similar surface chemistry-Xbridge C(18) and Atlantis d18 were tested. When using UPLC, the signal-to-noise ratio could be improved by a factor of 2-50 for different folate derivatives and the run time could be reduced fourfold without sacrificing separation efficiency. The applicability of UPLC to real food samples was demonstrated

Keywords

Ultra performance liquid chromatography; Ultra fast separation; Sub-2 mu m particles; Folate food analysis LIQUID-CHROMATOGRAPHY; MASS-SPECTROMETRY; HIGH-TEMPERATURE; HIGH-PRESSURE; DERIVATIVES; STABILITY; BERRIES; FINLAND; ASSAY; ACID

Published in

Chromatographia
2011, Volume: 73, number: 3-4, pages: 219-225