Lilliehöök, Inger
- Institutionen för kliniska vetenskaper, Sveriges lantbruksuniversitet
Forskningsartikel2009Vetenskapligt granskad
Lilliehöök, Inger; Tvedten, Harold
The Sysmex XT-2000iV performed as well as the CELL-DYN on blood samples from dogs, cats, and horses with a variety of hematologic abnormalities. In addition, the Sysmex detected large platelets and provided accurate reticulocyte counts.
Blood cell; hematology analyzer; laboratory; instrumentation; validation
Veterinary Clinical Pathology
2009, volym: 38, nummer: 2, sidor: 163-174
Utgivare: AMER SOC VETERINARY CLINICAL PATHOLOGY
Veterinärmedicin
Husdjursvetenskap
https://res.slu.se/id/publ/49839